Chapter 5 Theory of Operation<br />

Backplane<br />

Backplane<br />

Unless otherwise noted, components in this discussion are located on the<br />

A3 circuit assembly (34970-66503).<br />

The backplane contains three connectors, P101, P102, and P103 for<br />

connection to the plug-in modules. The parallel lines in these connectors<br />

are divided into two groups to form the analog bus and digital bus.<br />

Analog Bus<br />

The analog bus connects the signals from the plug-in modules to the<br />

Internal DMM. There are five lines in the analog bus, HI, LO,<br />

OHMS_HI, OHMS_LO, and AMPS. The HI and LO lines are protected<br />

from overvoltages by E101, E102, RV101, RV102, R101, R160, L101,<br />

L102, and C109.<br />

P105 makes the analog bus connection to the internal DMM.<br />

Digital Bus<br />

The digital bus uses 10 lines for communication and control. P104 makes<br />

the digital bus connection to the earth referenced logic and floating logic.<br />

Signal<br />

P101, P102, P103<br />

Pins<br />

Comments<br />

Slot ID A6, A7, A8 Unique binary code for each slot.<br />

DATA_IN B6 Serial module data from the floating logic.<br />

DATA_OUT C6 Serial module data to the floating logic.<br />

DGND A5, B5, B7, C5, C7 Earth referenced digital ground.<br />

+5 V A3, B3, C3 Earth referenced module power supply.<br />

Earth Ground A1, B1, C1 Earth referenced zap return ground.<br />

RST B8 Module reset from the earth referenced logic.<br />

SRQ C8 Module service request to the earth<br />

referenced logic.<br />