01.02.2014 Views

Agilent 34970A/34972A Data Acquisition / Switch Unit

Agilent 34970A/34972A Data Acquisition / Switch Unit

Agilent 34970A/34972A Data Acquisition / Switch Unit

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Chapter 4 Calibration Procedures<br />

Test Considerations<br />

Test Considerations<br />

To ensure proper instrument operation, verify that you have selected the<br />

correct power line voltage prior to attempting any procedure in this<br />

chapter. See “If the Instrument Does Not Turn On”, on page 50.<br />

Errors may be induced by AC signals present on the input leads during a<br />

self-test. Long test leads can also act as an antenna causing pick-up of<br />

AC signals.<br />

For optimum performance, all procedures should comply with the<br />

following recommendations:<br />

• Assure that the calibration ambient temperature is stable and<br />

between 18 °C and 28 °C. Ideally the calibration should be performed<br />

at 23 °C ±1 °C.<br />

• Assure ambient relative humidity is less than 80%.<br />

• Allow a 2-hour warm-up period with a copper short connected and the<br />

multiplexer module installed before verification or adjustment.The<br />

connections are shown in the figure on page 76.<br />

• Use shielded twisted pair PTFE insulated cables to reduce settling<br />

and noise errors. Keep the input cables as short as possible.<br />

• Connect the input cable shield to the source LO output.Except where<br />

noted in the procedures, connect the calibrator LO source to earth<br />

ground.<br />

Two-wire Ohms measurements are affected by the entire path length,<br />

including the plug-in card trace length and slot trace lengths. On the<br />

34901A, Channel 10 is recommended as the median path length (on the<br />

34902A, use Channel 8) for 2-wire Ohms verification and adjustments.<br />

Install the input multiplexer in slot 200.<br />

Because the instrument is capable of making highly accurate<br />

measurements, you must take special care to ensure that the calibration<br />

standards and test procedures used do not introduce additional errors.<br />

Ideally, the standards used to verify and adjust the instrument should be<br />

an order of magnitude more accurate than each instrument range full<br />

scale error specification.<br />

For the DC voltage, DC current, and resistance gain verification<br />

measurements, you should take care to ensure the calibrator’s “0” output<br />

is correct. If necessary, the measurements can be referenced to the<br />

calibrator’s “0” output using Mx + B scaling (see page 64). You will need<br />

to set the offset for each range of the measuring function being verified.<br />

4<br />

81

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!