Chapter 4 Calibration Procedures<br />
Relay Verification<br />
...Continued from previous page<br />
External DMM Ohmmeter Connections<br />
Test<br />
#<br />
Channel<br />
Closed*<br />
HI LO HI<br />
Sense<br />
LO<br />
Sense<br />
Measured Value<br />
Test<br />
Limit<br />
Relay<br />
Measured<br />
17 Ch 31 & 41 P3 P3 P4 P4 ________Ohms 2.00 K301<br />
18 Ch 32 & 42 P3 P3 P4 P4 ________Ohms 2.00 K302<br />
19 Ch 33 & 43 P3 P3 P4 P4 ________Ohms 2.00 K303<br />
20 Ch 34 & 44 P3 P3 P4 P4 ________Ohms 2.00 K304<br />
21 Ch 35 & 45 P3 P3 P4 P4 ________Ohms 2.00 K305<br />
22 Ch 36 & 46 P3 P3 P4 P4 ________Ohms 2.00 K306<br />
23 Ch 37 & 47 P3 P3 P4 P4 ________Ohms 2.00 K307<br />
24 Ch 38 & 48 P3 P3 P4 P4 ________Ohms 2.00 K308<br />
25 Ch 41 & 11 P4 P4 P1 P1 ________Ohms 2.00 K401<br />
26 Ch 42 & 12 P4 P4 P1 P1 ________Ohms 2.00 K402<br />
27 Ch 43 & 13 P4 P4 P1 P1 ________Ohms 2.00 K403<br />
28 Ch 44 & 14 P4 P4 P1 P1 ________Ohms 2.00 K404<br />
29 Ch 45 & 15 P4 P4 P1 P1 ________Ohms 2.00 K405<br />
30 Ch 46 & 16 P4 P4 P1 P1 ________Ohms 2.00 K406<br />
31 Ch 47 & 17 P4 P4 P1 P1 ________Ohms 2.00 K407<br />
32 Ch 48 & 18 P4 P4 P1 P1 ________Ohms 2.00 K408<br />
*Only the channel currently under test should be closed at one time. All other channels should be open.<br />
114