Chapter 6 Service<br />

Self-Test Procedures<br />

Self-Tests<br />

A complete self-test performs the following tests. A failing test is<br />

indicated by the test number and description in the display. For a<br />

complete list of error messages, see Chapter 5 in the <strong>Agilent</strong> <strong>34970A</strong>/<br />

<strong>34972A</strong> User’s Guide.<br />

601 Front panel not responding The main CPU A1U205 attempts to<br />

establish serial communications with the front panel processor A2U1.<br />

During this test, A2U1 turns on all display segments. Communication<br />

must function in both directions for this test to pass. If this error is<br />

detected during power-up self-test, the instrument will beep. This error<br />

is only readable from the remote interface.<br />

602 RAM read/write failed This test writes and reads a 55 and AA<br />

checkerboard pattern to each address of RAM. Any incorrect readback<br />

will cause a test failure. This error is only readable from the remote<br />

interface.<br />

603 A / D sync stuck The main CPU issues an A/ D sync pulse to A1U209<br />

and A1U205 to latch the value in the ADC slope counters. A failure is<br />

detected when a sync interrupt is not recognized and a subsequent timeout<br />

occurs.<br />

604 A / D slope convergence failed The input amplifier is configured to<br />

the measure zero (MZ) state in the 10 V range. This test checks whether<br />

the ADC integrator produces nominally the same number of positive and<br />

negative slope decisions (± 10%) during a 20 ms interval.<br />

605 Cannot calibrate rundown gain This test checks the nominal gain<br />

between the integrating ADC and the A1U205 on-chip ADC. This error is<br />

reported if the procedure can not run to completion due to a hardware<br />

failure.<br />

606 Rundown gain out of range This test checks the nominal gain<br />

between the integrating ADC and the A1U205 on-chip ADC. The<br />

nominal gain is check to ± 10% tolerance.<br />

178