Chapter 4 Calibration Procedures<br />
Relay Verification<br />
Tests 1 - 32:<br />
Open all channels on the module by performing a Factory Reset (press<br />
and select “Recall State”; press again and select “Factory<br />
Reset”). For each test, close only the channels shown in the “Channels<br />
Closed” column below (module in slot 200). Record the 4-wire ohms<br />
measurements from the external DMM in the following table.<br />
Note: To prepare the module between tests, press and hold down until<br />
“Card Reset” is displayed on the front panel.<br />
External DMM Ohmmeter Connections<br />
Module Reference<br />
Test<br />
#<br />
Channel<br />
Closed*<br />
HI LO HI<br />
Sense<br />
LO<br />
Sense<br />
Measured Value<br />
Test<br />
Limit<br />
Relay<br />
Measured<br />
1 Ch 11 & 41 P1 P1 P4 P4 ________Ohms 2.00 K101<br />
2 Ch 12 & 42 P1 P1 P4 P4 ________Ohms 2.00 K102<br />
4<br />
3 Ch 13 & 43 P1 P1 P4 P4 ________Ohms 2.00 K103<br />
4 Ch 14 & 44 P1 P1 P4 P4 ________Ohms 2.00 K104<br />
5 Ch 15 & 45 P1 P1 P4 P4 ________Ohms 2.00 K105<br />
6 Ch 16 & 46 P1 P1 P4 P4 ________Ohms 2.00 K106<br />
7 Ch 17 & 47 P1 P1 P4 P4 ________Ohms 2.00 K107<br />
8 Ch 18 & 48 P1 P1 P4 P4 ________Ohms 2.00 K108<br />
9 Ch 21 & 41 P2 P2 P4 P4 ________Ohms 2.00 K201<br />
10 Ch 22 & 42 P2 P2 P4 P4 ________Ohms 2.00 K202<br />
11 Ch 23 & 43 P2 P2 P4 P4 ________Ohms 2.00 K203<br />
12 Ch 24 & 44 P2 P2 P4 P4 ________Ohms 2.00 K204<br />
13 Ch 25 & 45 P2 P2 P4 P4 ________Ohms 2.00 K205<br />
14 Ch 26 & 46 P2 P2 P4 P4 ________Ohms 2.00 K206<br />
15 Ch 27 & 47 P2 P2 P4 P4 ________Ohms 2.00 K207<br />
16 Ch 28 & 48 P2 P2 P4 P4 ________Ohms 2.00 K208<br />
*Only the channel currently under test should be closed at one time. All other channels should be open.<br />
Continued on next page...<br />