Chapter 4 Calibration Procedures<br />
Internal DMM Verification Tests<br />
AC VOLTS Gain Verification Test<br />
Configuration: AC Volts<br />
LF 3 HZ:SLOW (in the Advanced menu)<br />
1 Make sure you have read “Test Considerations” on page 81.<br />
2 Select Channel 210, set the AC VOLTS function and the 3 Hz input<br />
filter. With the slow filter selected, each measurement takes 7 seconds to<br />
complete. Before executing each test, you must press to enable<br />
reading monitoring on the selected channel (or use the ROUTe:MON<br />
command from the remote interface).<br />
3 Select each range in the order shown below. Compare measurement<br />
results to the appropriate test limits shown in the table. (Be certain to<br />
allow for appropriate source settling.)<br />
Input Range Quick<br />
Error from Nominal<br />
Check<br />
V rms Frequency 24 hour 90 day 1 year<br />
100 mV<br />
100 mV<br />
± 70 V ± 90 V ± 100 V<br />
100 mV<br />
1 V<br />
1 V<br />
10 V<br />
10 V<br />
10 V<br />
10 mV [1]<br />
100 V<br />
100 V<br />
300 V<br />
300 V [2] 1 kHz<br />
50 kHz<br />
1 kHz<br />
50 kHz<br />
1 kHz<br />
50 kHz<br />
10 Hz<br />
1 kHz<br />
1 kHz<br />
50 kHz<br />
1 kHz<br />
50 kHz<br />
Q<br />
Q<br />
Q<br />
± 150 V<br />
± 700 V<br />
± 1.5 mV<br />
± 7 mV<br />
± 15 mV<br />
± 7 mV<br />
± 34 V<br />
± 70 mV<br />
± 150 mV<br />
± 270 mV<br />
± 600 mV<br />
± 160 V<br />
± 900 V<br />
± 1.6 mV<br />
± 9 mV<br />
± 16 mV<br />
± 9 mV<br />
± 45 V<br />
± 90 mV<br />
± 160 mV<br />
± 390 mV<br />
± 690 mV<br />
± 170 V<br />
± 1 mV<br />
± 1.7 mV<br />
± 10 mV<br />
± 17 mV<br />
± 10 mV<br />
± 46 V<br />
± 100 mV<br />
± 170 mV<br />
± 420 mV<br />
± 720 mV<br />
1 V<br />
10 V<br />
100 mV<br />
100 v<br />
300V<br />
[1] For this test, isolate the calibrator’s output from earth ground.<br />
[2] Some calibrators may have difficulty driving the internal DMM and cable load at this V-Hz<br />
output. Use short, low capacitance cable to reduce calibration loading. Verification can be<br />
performed at >195 Vrms. New test limits can be computed from the accuracy specification<br />
shown in Chapter 1 for the actual test conditions used.<br />
Q: Quick performance verification test points.<br />
Note: The 50 kHz AC voltage test points may fail performance<br />
verification if the internal shields have been removed and reinstalled. See<br />
“Gain Adjustment,” on page 92, for further information on how to<br />
recalibrate the AC voltage function.<br />
88