Chapter 4 Calibration Procedures<br />

Internal DMM Verification Tests<br />

AC VOLTS Gain Verification Test<br />

Configuration: AC Volts<br />

LF 3 HZ:SLOW (in the Advanced menu)<br />

1 Make sure you have read “Test Considerations” on page 81.<br />

2 Select Channel 210, set the AC VOLTS function and the 3 Hz input<br />

filter. With the slow filter selected, each measurement takes 7 seconds to<br />

complete. Before executing each test, you must press to enable<br />

reading monitoring on the selected channel (or use the ROUTe:MON<br />

command from the remote interface).<br />

3 Select each range in the order shown below. Compare measurement<br />

results to the appropriate test limits shown in the table. (Be certain to<br />

allow for appropriate source settling.)<br />

Input Range Quick<br />

Error from Nominal<br />

Check<br />

V rms Frequency 24 hour 90 day 1 year<br />

100 mV<br />

100 mV<br />

± 70 V ± 90 V ± 100 V<br />

100 mV<br />

1 V<br />

1 V<br />

10 V<br />

10 V<br />

10 V<br />

10 mV [1]<br />

100 V<br />

100 V<br />

300 V<br />

300 V [2] 1 kHz<br />

50 kHz<br />

1 kHz<br />

50 kHz<br />

1 kHz<br />

50 kHz<br />

10 Hz<br />

1 kHz<br />

1 kHz<br />

50 kHz<br />

1 kHz<br />

50 kHz<br />

Q<br />

Q<br />

Q<br />

± 150 V<br />

± 700 V<br />

± 1.5 mV<br />

± 7 mV<br />

± 15 mV<br />

± 7 mV<br />

± 34 V<br />

± 70 mV<br />

± 150 mV<br />

± 270 mV<br />

± 600 mV<br />

± 160 V<br />

± 900 V<br />

± 1.6 mV<br />

± 9 mV<br />

± 16 mV<br />

± 9 mV<br />

± 45 V<br />

± 90 mV<br />

± 160 mV<br />

± 390 mV<br />

± 690 mV<br />

± 170 V<br />

± 1 mV<br />

± 1.7 mV<br />

± 10 mV<br />

± 17 mV<br />

± 10 mV<br />

± 46 V<br />

± 100 mV<br />

± 170 mV<br />

± 420 mV<br />

± 720 mV<br />

1 V<br />

10 V<br />

100 mV<br />

100 v<br />

300V<br />

[1] For this test, isolate the calibrator’s output from earth ground.<br />

[2] Some calibrators may have difficulty driving the internal DMM and cable load at this V-Hz<br />

output. Use short, low capacitance cable to reduce calibration loading. Verification can be<br />

performed at >195 Vrms. New test limits can be computed from the accuracy specification<br />

shown in Chapter 1 for the actual test conditions used.<br />

Q: Quick performance verification test points.<br />

Note: The 50 kHz AC voltage test points may fail performance<br />

verification if the internal shields have been removed and reinstalled. See<br />

“Gain Adjustment,” on page 92, for further information on how to<br />

recalibrate the AC voltage function.<br />

88