Chapter 4 Calibration Procedures<br />

Calibration Security<br />

Calibration Security<br />

This feature allows you to enter a security code to prevent accidental or<br />

unauthorized adjustments of the instrument. When you first receive<br />

your instrument, it is secured. Before you can adjust the instrument, you<br />

must unsecure it by entering the correct security code. See page 60 in<br />

Chapter 3 for a procedure to enter the security code.<br />

• The security code is set to “HP034970” or “AT034972”, depending on<br />

the product number, when the instrument is shipped from the<br />

factory. The security code is stored in non-volatile memory, and does<br />

not change when power has been off, after a Factory Reset (*RST<br />

command), or after an Instrument Preset (SYSTem:PRESet<br />

command).<br />

• The security code may contain up to 12 alphanumeric characters. The<br />

first character must be a letter, but the remaining characters can be<br />

letters, numbers, or an underscore ( _ ). You do not have to use all 12<br />

characters but the first character must always be a letter.<br />

Note: If you forget your security code, you can disable the security feature<br />

by adding a jumper inside the instrument as described on the following<br />

page.<br />