Chapter 4 Calibration Procedures<br />
Calibration Security<br />
Calibration Security<br />
This feature allows you to enter a security code to prevent accidental or<br />
unauthorized adjustments of the instrument. When you first receive<br />
your instrument, it is secured. Before you can adjust the instrument, you<br />
must unsecure it by entering the correct security code. See page 60 in<br />
Chapter 3 for a procedure to enter the security code.<br />
• The security code is set to “HP034970” or “AT034972”, depending on<br />
the product number, when the instrument is shipped from the<br />
factory. The security code is stored in non-volatile memory, and does<br />
not change when power has been off, after a Factory Reset (*RST<br />
command), or after an Instrument Preset (SYSTem:PRESet<br />
command).<br />
• The security code may contain up to 12 alphanumeric characters. The<br />
first character must be a letter, but the remaining characters can be<br />
letters, numbers, or an underscore ( _ ). You do not have to use all 12<br />
characters but the first character must always be a letter.<br />
Note: If you forget your security code, you can disable the security feature<br />
by adding a jumper inside the instrument as described on the following<br />
page.<br />