Chapter 5 Theory of Operation<br />
<strong>Switch</strong> Modules<br />
In addition to the +5 Volt power supply (Vcc) and ground, the module<br />
controller uses four lines for control and communication:<br />
• RESET, from the Earth Referenced Logic A1U305. RESET is common<br />
to all three slots. The module controller performs a reset when this<br />
line goes high. Reset conditions vary for each plug-in.<br />
• SRQ, to the Earth Referenced Logic A1U305. The SRQ line is a<br />
wired-OR line that can be driven by any plug-in. Consequently, any<br />
module that asserts SRQ (line low), asserts this line in all other slots<br />
and at the Earth Reference Logic.<br />
• DATA IN, from the Floating Logic A1U205 via the opto-isolator<br />
A1U312. This line is connected in common to all three slots.<br />
• DATA OUT, from the module controller to the Floating Logic A1U205<br />
via the opto-isolator A1U213. This line is a wired-OR line that can be<br />
driven by any module.<br />
The DATA IN and DATA OUT lines are optically isolated from the<br />
floating logic controller. These lines communicate with the Floating<br />
Logic using an asynchronous serial bit stream.<br />
The serial communications use an 11 bit protocol; a start bit, 8 data bits,<br />
an attention bit, and a stop bit. The attention bit is 1 if the 8 data bits are<br />
an address/command, or 0 if the 8 data bits modify or provide data for<br />
the previously sent command.<br />
The module controller uses the hardwired slot-ID bits to decode the<br />
serial bit stream address. When the address/command message address<br />
matches the slot-ID, the plug-in is selected and responds to the following<br />
commands. All other plug-ins will ignore the commands until a new<br />
address/command message is received.<br />
A 12 MHz crystal, Y101 is the clock for the module controller. The<br />
module ID is a four bit pattern set through R106, R107, R108, and R109.<br />
The ferroelectric RAM U150 provides data storage of the relay usage on<br />
the module (see page 65) and the thermocouple reference junction<br />
temperature corrections (see page 122). <strong>Data</strong> in and out of U150 is serial.<br />
The 34901A, 34902A, and 34908A each have two onboard, solid state<br />
temperature sensors, U151 and U152, physically located near the<br />
isothermal block at the input connections. The temperature sensors are<br />
used as the thermocouple temperature reference.<br />