Chapter 4 Calibration Procedures<br />

Relay Verification<br />

(Optional)<br />

34904A Relay Contact Resistance Verification<br />

This optional procedure uses an external DMM to make 4-wire ohms<br />

measurements across the relay contacts on the 34904A.<br />

1 Be sure to read “Plug-in Module Test Considerations” on page 97.<br />

2 Connect coppers shorts from H to L on each of the eight columns (COL1<br />

through COL8) as shown below. Connect four dual banana plugs to the<br />

four rows as shown below (ROW1 through ROW4). Be sure to route your<br />

wiring for proper strain relief and install the module cover. Install the<br />

34904A in slot 200 of the <strong>34970A</strong>/<strong>34972A</strong>.<br />

Connections for 34904A Verification Tests<br />

112