Chapter 4 Calibration Procedures<br />

Relay Verification<br />

Test 37:<br />

Test<br />

#<br />

Channel<br />

Configured<br />

Install the 34901A in slot 200 of the <strong>34970A</strong>/<strong>34972A</strong>. Select and<br />

configure Channel 21 as follows: DC current, 1 amp range, and 5½ digits.<br />

Enable reading monitoring by pressing on the selected channel (or<br />

use the ROUTe:MON command). Record the 4-wire ohms measurement<br />

from the external DMM in the following table.<br />

External DMM Ohmmeter Connections<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

37 Ch 21 P4 P4 P5 P5 ________Ohms 2.00 K521<br />

Module Reference<br />

Tests 38 - 39:<br />

Open all channels on the module by performing a Factory Reset. Close<br />

Channel 21 (module in slot 200). Remove the 34901A from the <strong>34970A</strong>/<br />

<strong>34972A</strong> and do not reinstall it for the remaining tests. Using the external<br />

DMM, make a 4-wire ohms measurement between the L and I terminals<br />

on Channel 22. Record the measured value as Test 38 in the following<br />

table.<br />

4<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

Channel<br />

Closed<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured Value<br />

Test<br />

Limit<br />

Relay<br />

Measured<br />

38 Ch 21 Ch 22 I Ch 22 L Ch 22 I Ch 22 L ________Ohms — —<br />

39 Subtract (Test 38 - Test 5) ________Ohms 2.00 K524<br />

105