Chapter 4 Calibration Procedures<br />

Internal DMM Verification Tests<br />

Internal DMM Verification Tests<br />

These procedures use inputs connected to a 34901A 20-Channel<br />

Multiplexer (see page 76) installed in slot 200.<br />

Zero Offset Verification<br />

This procedure is used to check the zero offset performance of the<br />

internal DMM. Verification checks are only performed for those<br />

functions and ranges with unique offset calibration constants.<br />

Measurements are checked for each function and range as described in<br />

the procedure below.<br />

Zero Offset Verification Procedure<br />

1 Make sure you have read “Test Considerations” on page 81.<br />

2 This procedure will measure the shorts installed on Channels 209 and<br />

219. Leave the Amps input connections (Channel 221) open.<br />

4<br />

Continued on next page...<br />