Chapter 4 Calibration Procedures<br />
Recommended Test Equipment<br />
Recommended Test Equipment<br />
The test equipment recommended for the performance verification and<br />
adjustment procedures is listed below. If the exact instrument is not<br />
available, substitute calibration standards of equivalent accuracy.<br />
A suggested alternate method would be to use the <strong>Agilent</strong> 3458A<br />
8½-digit Digital Multimeter to measure less accurate yet stable sources.<br />
The output value measured from the source can be entered into the<br />
instrument as the target calibration value.<br />
Application Recommended Equipment Accuracy Requirements<br />
Zero Calibration [1]<br />
DC Voltage [1]<br />
Dc Current [1]<br />
Resistance [1] ]<br />
AC Voltage [1]<br />
AC Current [1]<br />
Frequency [1]<br />
Analog Output<br />
34907A<br />
Thermocouple<br />
Reference Junction<br />
34901A<br />
34902A<br />
34908A<br />
Relay contact resistance<br />
All switch modules<br />
None<br />
Fluke 5700A<br />
Fluke 5700A/5725A<br />
Fluke 5700A<br />
Fluke 5700A/5725A<br />
Fluke 5700A/5725A<br />
<strong>Agilent</strong> 33220A<br />
<strong>Agilent</strong> 34401A<br />
Thermistor YSI 44031 (two) [2]<br />
J Type Calibrated Thermocouple<br />
Triple Point Cell<br />
<strong>Agilent</strong> 34401A<br />
4 -terminal all copper short<br />