Chapter 4 Calibration Procedures<br />

Relay Verification<br />

(Optional)<br />

Tests 1 - 5:<br />

34901A Relay Contact Resistance Verification<br />

This optional procedure uses an external DMM to make 4-wire ohms<br />

measurements across the relay contacts on the 34901A. The measured<br />

resistance is the series resistance of the two relay contacts (both contacts<br />

are in the same relay).<br />

Note: Be sure to read “Plug-in Module Test Considerations” on page 97.<br />

See the diagram on page 100 for the required connections for each test<br />

(be sure to probe the components at the indicated location). For these<br />

measurements, the 34901A is not installed in the <strong>34970A</strong>/<strong>34972A</strong>.<br />

Record the<br />

4-wire ohms measurements from the external DMM in the table below.<br />

Note: The connections to the external DMM are different for each of Tests<br />

1, 2, 4, and 5. Be sure to verify the connections shown in the table below<br />

for each of the four tests.<br />

4<br />

Module Reference<br />

External DMM Ohmmeter Connections<br />

Test<br />

#<br />

HI LO HI<br />

Sense<br />

LO<br />

Sense<br />

Measured<br />

Value<br />

1 L401 J101, C14 L401 J101, C14 _____________ Ohms<br />

2 L402 J101, C15 L402 J101, C15 _____________ Ohms<br />

3 Add (Test 1 + Test 2) _____________ Ohms<br />

4 F501 C21 I F501 C21 I _____________ Ohms<br />

5 F502 C22 I F502 C22 I _____________ Ohms<br />

99