Chapter 4 Calibration Procedures<br />
Relay Verification<br />
(Optional)<br />
Tests 1 - 5:<br />
34901A Relay Contact Resistance Verification<br />
This optional procedure uses an external DMM to make 4-wire ohms<br />
measurements across the relay contacts on the 34901A. The measured<br />
resistance is the series resistance of the two relay contacts (both contacts<br />
are in the same relay).<br />
Note: Be sure to read “Plug-in Module Test Considerations” on page 97.<br />
See the diagram on page 100 for the required connections for each test<br />
(be sure to probe the components at the indicated location). For these<br />
measurements, the 34901A is not installed in the <strong>34970A</strong>/<strong>34972A</strong>.<br />
Record the<br />
4-wire ohms measurements from the external DMM in the table below.<br />
Note: The connections to the external DMM are different for each of Tests<br />
1, 2, 4, and 5. Be sure to verify the connections shown in the table below<br />
for each of the four tests.<br />
4<br />
Module Reference<br />
External DMM Ohmmeter Connections<br />
Test<br />
#<br />
HI LO HI<br />
Sense<br />
LO<br />
Sense<br />
Measured<br />
Value<br />
1 L401 J101, C14 L401 J101, C14 _____________ Ohms<br />
2 L402 J101, C15 L402 J101, C15 _____________ Ohms<br />
3 Add (Test 1 + Test 2) _____________ Ohms<br />
4 F501 C21 I F501 C21 I _____________ Ohms<br />
5 F502 C22 I F502 C22 I _____________ Ohms<br />
99