Chapter 5 Theory of Operation<br />

<strong>Switch</strong> Modules<br />

34901A<br />

Components in this discussion are located on the A1 circuit assembly<br />

(34901-66501).<br />

The control circuitry has four groupings of latches, relay drivers and<br />

relays. The three control lines (SEL_A, SEL_B, and SEL_C) from the<br />

module controller are divided into five control lines by the binary to octal<br />

converter U109.<br />

The row latch, U102, and column latches, U103 and U104, control the<br />

relays. The row drivers are divided into four groups of set and reset<br />

drivers. Each group of row drivers controls five relays. The column<br />

drivers operate as a pair. There are ten column drivers each controlling<br />

two relays. The analog bus backplane relays are non-latching.<br />