Chapter 5 Theory of Operation<br />
Internal DMM<br />
Ohms Current Source<br />
Unless otherwise noted, components in this discussion are located on the<br />
A4 circuit assembly (34970-66504).<br />
The ohms current source flows from the Input HI terminal to the Input<br />
LO terminal for both the 2-wire and 4-wire ohms functions. Each current<br />
value is generated by forcing a stable, precise voltage across a stable<br />
resistance. The value of the current becomes part of the range gain<br />
constant stored during calibration.<br />
4<br />
The +7 V reference voltage is used to generate a stable reference current<br />
with U201A. R201 and R202 are the resistance references for the current<br />
sources as shown in the table below. The IREF current is used to produce<br />
a precise voltage drop across the 28.57 k resistor in U102D-4. The<br />
IREF generated using R202 produces an approximate 5 V drop across<br />
the 28.57 k resistor. The IREF generated using R201 produces an<br />
approximate 0.5 V drop. This voltage is used to force a reference voltage<br />
across the selected current source range resistor (5 k, 50 k, 500 k,<br />
1M) by U201B. The resulting precision current flows through JFET<br />
Q202 and protection circuit Q203 to Q211, and CR202 to relay K102<br />
where it is switched to the Input HI terminal for ohms measurements.<br />
The protection circuits are designed to protect the ohms current source<br />
from inadvertently applied voltages in excess of ±1000 V. Protection from<br />
large positive voltages is provided by the reverse breakdown voltage of<br />
CR202. Protection from large negative voltages is provided by the sum of<br />
the collector to base breakdown voltages of Q203, Q205, Q207, and Q209.<br />
Bias for these transistors is provided by Q211 and R203 to R206 while<br />
negative over voltages are applied.<br />
5<br />
Ohms Range Current Open Circuit<br />
Voltage<br />
100 1 ma<br />
9 V<br />
100 M [1] 500 nA [1] 5 V<br />
1 k 1 mA<br />
9 V<br />
10 k 100 A 9 V<br />
100 k 10 A<br />
9 V<br />
1 M 5 A<br />
9 V<br />
10 M 500 nA 14 V<br />
Compliance<br />
Limit<br />
2.5 V<br />
2.5 V<br />
4 V<br />
4V<br />
8 V<br />
10 V<br />
Reference<br />
R202<br />
R202<br />
R202<br />
R202<br />
R202<br />
R201<br />
R201<br />
Isource R<br />
U102D<br />
5 k<br />
5 k<br />
50 k<br />
500 k<br />
1 M<br />
1 M<br />
1 M<br />
[1] Measured in parallel with the internal 10 M resistor.<br />
143