Chapter 6 Service<br />

Self-Test Procedures<br />

619 Ohms 1 mA source failed This test configures the 10 V range with the<br />

internal 1 0 M 100:1 divider A4U102 connected across the input. The 1<br />

mA current source is connected. A 20 ms ADC measurement is<br />

performed and the result is checked against a limit of 7 V ± 3.5 V.<br />

620 AC rms zero failed This test configures to the 100 mV AC range with<br />

the AC input grounded through A4K103. The internal residual noise of<br />

the AC section is measured and checked against a limit of –10 mV to 70<br />

mV at the output of the rms-to-dc converter.<br />

621 AC rms full scale failed This test configures for the 100 mV AC range.<br />

The 1 mA ohms current source is switched on the charge the AC input<br />

capacitor A4C301. This produces a pulse on the output of the rms-to-dc<br />

converter which is sampled 100 ms after the current is applied. A 20 ms<br />

A/D measurement is performed and checked against a limit of 10 V ± 8.5<br />

V into the ADC.<br />

622 Frequency counter failed This test configures for the 100 mV AC<br />

range. This test immediately follows test 621. With A4C301 holding<br />

charge from test 621 the AC input is now switched to ground through<br />

A4K103. This produces a positive pulse on the input to the frequency<br />

comparator A4U310. While C301 discharges, the ENAB FREQ bit is<br />

toggled four times to produce a frequency input to the counter logic in<br />

A1U205. A failure occurs if the counter can not measure the frequency<br />

input.<br />

4<br />

623 Cannot calibrate precharge This test configures to the 100 V DC<br />

range with no input. The ADC is configured for 200 ms measurements.<br />

The A1U205 pulse width modulated (PWM) DAC output (C224) is set to<br />

about 4 volts. A reading is taken in with A4U101 in the MC state. A<br />

second reading is taken in the PRE state. The precharge amplifier<br />

voltage offset is calculated. The A1U205 DAC output is set to about 1.5<br />

volts and the precharge offset is measured again. The gain of the offset<br />

adjustment is calculated. This test assures a precharge amplifier offset is<br />

achievable.<br />

6<br />

181