Chapter 4 Calibration Procedures<br />
Calibration Message<br />
Calibration Message<br />
The instrument allows you to store one message in calibration memory.<br />
For example, you can store such information as the date when the last<br />
calibration was performed, the date when the next calibration is due, the<br />
instrument’s serial number, or even the name and phone number of the<br />
person to contact for a new calibration.<br />
• You can record a calibration message only from the remote interface<br />
and only when the instrument is unsecured. You can read the<br />
message from either the front-panel or over the remote interface. You<br />
can read the calibration message whether the instrument is secured<br />
or unsecured.<br />
• The calibration message may contain up to 40 characters. From the<br />
front panel, you can view 13 characters of the message at a time.<br />
Press to scroll through the text of the message. Press again to<br />
increase the scrolling speed.<br />
4<br />
Calibration Count<br />
You can query the instrument to determine how many calibrations have<br />
been performed. Note that your instrument was calibrated before it left<br />
the factory. When you receive your instrument, be sure to read the count<br />
to determine its initial value.<br />
• The calibration count increments up to a maximum of 65,535 after<br />
which it rolls over to “0”. Since the value increments by one for each<br />
calibration point, a complete calibration may increase the value by<br />
many counts.<br />
• The calibration count is also incremented with calibrations of the<br />
DAC channels on the multifunction module.<br />
• Front-Panel Operation:<br />
CAL COUNT<br />
• Remote Interface Operation:<br />
CALibration:COUNt?<br />
79