Chapter 5 Theory of Operation<br />

<strong>Switch</strong> Modules<br />

U101 controls the relays on the module using an 8-bit data bus and three<br />

control lines. The data lines are latched and applied to the relay drivers.<br />

U101 enters a low-power idle mode when inactive. U101 responds when<br />

a command is received or when a scheduled reference junction<br />

temperature measurement is taken.<br />

The relays use a buffered +5 Volt power supply. U101 supplies two drive<br />

enable lines (DR_EN and +5NL_EN) that connect Vcc from the digital<br />

bus with the relay drive lines through Q101 (+5R or +5NL). To minimize<br />

the current through DGND caused by static discharge, the ground return<br />

(ZGND) is isolated from the backplane ground through a bead L102.<br />

Relay Drivers<br />

Two types of single-coil relays are used on the switch modules: latching<br />

and non-latching. Typical driver configurations are shown below.<br />