2.14 Operation with the Quad MAP Controller<br />

The Quad MAP Controller (QMC) will synchronize the operation of two, three,<br />

or four MAP systems together with their associated <strong>CinePlex</strong> systems.<br />

Startup<br />

The startup sequence is extremely important. RASPUTIN must be<br />

started before Studio is started so that the client registration process<br />

works properly. Otherwise, when a recording is eventually started by<br />

RASPUTIN neither the file name string nor the recording time 0 messages<br />

are received.<br />

If RASPUTIN exits or is closed for any reason, it should be brought<br />

back up, then Studio should be closed and restarted, as well.<br />

Procedure for Quad MAP Controller Operation<br />

1 Each of the MAP/<strong>CinePlex</strong> combination systems should be tested and<br />

validated to be operational in stand-alone mode before attempting to connect<br />

them for synchronized use with the QMC. If this has not been done for each of<br />

the systems, please do so now. Then power all components of all systems<br />

down.<br />

2 Select one MAP system to be the “Master” MAP system. Signals from the<br />

master MAP system will control all of the connected CPX systems and will<br />

provide power to the Quad MAP Controller.<br />

3 Configure the TIM board of the master MAP system to supply power to the<br />

Quad MAP Controller. Apply a jumper at the position marked “17” and “+5V”<br />

on jumper block W3 as shown.<br />

Version 3.0 107