Note that applying this correction can shift Markers to have a negative timestamp<br />

near the beginning of files. When this happens, the timestamps will be assigned<br />

the value of 0.<br />

In addition to the correcting for the camera latency, an independent correction for<br />

the tracking latency can be applied, as discussed in the next section.<br />

Note:<br />

Note:<br />

Note:<br />

As of <strong>CinePlex</strong> Editor version 1.2.0, these latency corrections are now enabled<br />

by default, with a default correction of 21 milliseconds. If data have been analyzed<br />

using previous versions of <strong>CinePlex</strong> Editor, the timestamp shifts after<br />

upgrading to version 1.2.0. To return to pre-version-1.2.0 behavior, one can disable<br />

the corrections by unchecking the checkboxes.<br />

When <strong>CinePlex</strong> is being used with <strong>Plexon</strong>’s Recorder system, there was an<br />

issue with Recorder versions prior to 2.3.2 that could artificially increase this<br />

latency by up to 80 milliseconds. If using <strong>CinePlex</strong> with Recorder, please<br />

upgrade the version of Recorder to version 2.3.2 or later. Also, use a video<br />

latency correction of 33 milliseconds as discussed above.<br />

When <strong>CinePlex</strong> is being used with <strong>Plexon</strong>’s MAP system, leaving the MAP on<br />

overnight could result in excessive latency - much greater than 30 milliseconds.<br />

If the MAP has been left on overnight, it is better to reset the MAP system.<br />