toggle switch above these connectors is in the down position to select the<br />

remote clock.<br />

6 Optionally, digital input signals from an external system can be input to the<br />

“DIGITAL INPUT” socket on the QMC where they are made available on the<br />

“DIGITAL OUTPUT” sockets to all MAP boxes.<br />

7 Trigger cable(s) from the AccesIO tail board to the camera(s).<br />

8 FireWire cable(s) from the FireWire card in the computer to the camera(s). If<br />

there are two cameras, there must be an empty FireWire socket between the<br />

two connections to the FireWire board or maximum frame rates will not be<br />

achievable.<br />

Cables Too Short between Camera(s) and MAP/<strong>CinePlex</strong> Computer<br />

<strong>Plexon</strong> supplies several lengths of these cables from 7 feet (2 meters) to 26<br />

feet (8 meters). Please contact <strong>Plexon</strong> if one of these standard lengths will<br />

work in the installation. Note that cable lengths over 26 feet result in intermittent<br />

video signals at higher frame rates and are NOT RECOMMENDED.<br />

Unstable Camera Mounting<br />

Camera mounts must be stable or the resulting video data will shake and not<br />

be repeatable. This can be the case, for example, when a tripod is used to<br />

mount a camera, instead of the preferred wall or ceiling mount. Please refer to<br />

“Appendix C- Lens Selection, Experiment Design, and Camera Installation”.<br />