Software Reference 168Titer TabUse the Titer tab on the Assay Analysis Window to specify the wellsin the plate that contains the standard to be used as the calibrationstandard. The Titer is used for calibration instead of the ladder.(Titer is not available for Glycan assays.)Figure 64. Assay Analysis Window - Titer TabThe Titer tab contains the following options and buttons:Titer Table - Lists the standard size, well locations, andconcentrations for the titer on the plate.• Size - Specifies the size of the standard in BP for DNA assays,nt for RNA assays, or kDa for Protein assays.• Window (%) - Specifies the tolerance window as a percent ofthe size for the standard to allow for small variations in size.• Found in Well- Specifies the location of the standard in theplate. (Not displayed for imported titers.)• Normalized Area - Specifies the titer peak normalized area forimported titers. The table for imported titers is Read Only andcannot be edited. (Not displayed for manually entered titers.)P/N 127170 Rev. 00 <strong>LabChip</strong> <strong>GX</strong> <strong>User</strong> <strong>Manual</strong> Caliper Life Sciences, Inc.