Data Analysis 66Selecting a Default Ladder (Continued)Figure 25. Default Ladder Settings5 The Ladder Peak Time and Area settings can be changed in theDefault Ladder table, but default ladder peaks cannot be addedor deleted and the ladder peak sizes cannot be changed.6 The Marker Time, Height and Area can be changed in theMarkers table.7 Click the Apply button at the bottom of the window to save thechanges to the active plate. Click the Apply Global button toapply the default ladder settings to all open data files in thecollection. (This allows you to import the default ladder into aplate that does not include any good ladders on the plate.)8 See “Using the Default Ladder for Alignment” on page 67 toalign wells in the plate using the default ladder.P/N 127170 Rev. 00 <strong>LabChip</strong> <strong>GX</strong> <strong>User</strong> <strong>Manual</strong> Caliper Life Sciences, Inc.