Software Reference 172Advanced Tab (Continued)The default values for the Concentration Size Correction weredetermined empirically by measuring a wide range of proteins ofknown size and concentration. This correction can be bypassed bysetting Amplitude = 1.0, Power = 0.0.Peak Integration ModeSpecifies the mode for determining the baseline.• Peak Baseline - A global peak baseline is used as the baselinefor all peaks. This baseline is determined by stitching togetherregions of low variance across the data signal.• Valley to Valley - Each peak is assigned a baseline by drawinga line beneath the peak joining the start and end of the peak atthe data signal values.P/N 127170 Rev. 00 <strong>LabChip</strong> <strong>GX</strong> <strong>User</strong> <strong>Manual</strong> Caliper Life Sciences, Inc.