Data Analysis 68Changing the Time Window for AnalysisThe Start Time and End Time parameters in the Peak Find tabdefine the time window within which peaks are found.To change the Start Time and End Time parameters for all wells inthe open assay:1 Select Analysis Analysis Settings to open the AssayAnalysis Window, and then click the Peak Find Tab.2 Change the parameters as necessary.3 Click the OK button to save the setting, reanalyze the data, andclose the window.Click the Apply button to apply the changes and reanalyze thedata, but keep the Assay Analysis window open.Aligning or Unaligning the Marker PeaksTo perform data analysis for DNA, Protein, RNA, and Glycanassays, the <strong>LabChip</strong> <strong>GX</strong> software aligns marker peaks included inthe sample wells with markers from the ladder.• To view the raw data without analysis, choose Analysis TurnOff Analysis.• To re-enable analysis, choose Analysis Turn On Analysis.The default setting for this function is enabled. Turning off theanalysis displays the data without aligning the markers in the wellsand the ladders.Figure 26. Data Before and After AlignmentP/N 127170 Rev. 00 <strong>LabChip</strong> <strong>GX</strong> <strong>User</strong> <strong>Manual</strong> Caliper Life Sciences, Inc.