Operation 33Adding a New Plate (Continued)• The instrument will move to five wells in each corner of theplate, punching holes in the ParaFilm.• The Punch test is acceptable if the punched hole is >0.5mmfrom the edge of the well.• Holes will never be perfectly centered.Possible sources of errors:• Sipper splay• Plate tolerances• Robot alignment error• If the holes are too close to the edge (