<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>5.16 Changing Main Power Fuses1. Turn off the main power switch.HIGH VOLTAGE—Disconnect the main power cord from its source andalso from the rear panel of the <strong>ICS</strong>-<strong>1000</strong>.HAUTE TENSION—Débranchez le cordon d'alimentation principal desa source et du panneau arrière du <strong>ICS</strong>-<strong>1000</strong>.HOCHSPANNUNG—Ziehen Sie das Netzkabel aus der Steckdose undder Netzbuchse auf der Rückseite des <strong>ICS</strong>-<strong>1000</strong>.2. The fuse holder is part of the main power receptacle (see Figure 5-20) on the<strong>ICS</strong>-<strong>1000</strong> rear panel. To remove the fuse holder, squeeze the tab on the top ofthe holder to release it and pull the holder straight out of its compartment.3. Replace the two fuses in the holder with new IEC 127 fast-blow fuses rated3.15 amps (P/N 954745). Dionex recommends always replacing both fuses.4. Reinsert the fuse holder into its compartment and push in until the tab clicksin place.5. Reconnect the main power cord and turn on the power.Figure 5-20. Main Power Fuse Holder5-30 Doc. 031879-01 3/03