<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>B.14.2 Priming with the Prime ButtonNOTEPerform this procedure after priming the eluentlines (see Section B.14.1).1. Check that the priming valve on the primary pump head is closed (seeFigure B-37).SecondaryPump HeadWaste Valve(open)Priming Valve(close)PrimaryPump HeadFigure B-37. Priming the Pump2. Open the waste valve on the secondary pump head by turning theknob one-quarter to one-half turn counterclockwise (seeFigure B-37). Opening the valve directs the eluent flow path to wasteand eliminates backpressure.3. Click Prime on the Chromeleon Control Panel (see Figure B-36). Thepump will begin pumping at approximately 3.0 mL/min.4. Continue priming the <strong>ICS</strong>-<strong>1000</strong> until no air bubbles are exiting thepump waste line.5. Press Pump Off.6. Close the waste valve. Do not overtighten. The pump is now readyfor operation.NOTEIf the two standard priming procedures described aboveare unsuccessful, prime the pump with isopropylalcohol.See Section 5.15 for instructions.B-38 Doc. 031879-01 3/03