<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>4.5 No Flow• Pump waste or priming valve open (see Figure 4-2).Close the valves by turning the knobs clockwise until fingertight. DO NOTOVERTIGHTEN! Overtightening may damage the valve and the pumphead.Waste ValvePriming ValveFigure 4-2. Waste and Priming Valves• Flow rate is set to 0Set the flow rate from either the Chromeleon Control Panel.• Eluent valve is closedOpen the eluent valve from either the Chromeleon Control Panel.• Pump not primedPrime the pump (see Section B.12).• Broken pump pistonReplace the piston (see Section 5.7).4-14 Doc. 031879-01 3/03