ICS-1000 Ion Chromatography System Operator's Manual

ICS-1000 Ion Chromatography System Operator's Manual

ICS-1000 Ion Chromatography System Operator's Manual

SHOW MORE
SHOW LESS
  • No tags were found...

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

2 • Features2.4 <strong>System</strong> Component DetailsThis section provides details about <strong>ICS</strong>-<strong>1000</strong> system components, including thevacuum degas assembly (optional), eluent valve, pump, injection valve, columnheater (optional), suppressor, and conductivity cell.2.4.1 Vacuum Degas Assembly (Optional)The vacuum degas assembly provides online eluent degassing at the timeand duration specified by the user. The assembly, which must be installedin the <strong>ICS</strong>-<strong>1000</strong> at the factory, consists of:• A single-channel degas chamber (with degas membranes) withinternal capacity of 17 mL• A dual-stage diaphragm vacuum pump• A solenoid valve• An on-board vacuum sensor• The electronics required to operate the vacuum pump• Tubing, fittings, and other accessoriesBy default, the <strong>ICS</strong>-<strong>1000</strong> monitors the degas pressure reading and turnsthe degas pump on and off as needed. A different degas operating modecan be selected from Chromeleon. To select degas operating options fromChromeleon, open the Server Configuration program, select the <strong>ICS</strong>-<strong>1000</strong>Doc. 031879-01 3/03 2-19