<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>5. Press Pump Off.6. Close the waste valve. Do not overtighten. The pump is now ready foroperation.3.7 Set <strong>System</strong> Operating ConditionsNOTEThis section is an overview of the steps needed to startup the system and begin running samples. Actualoperating parameters (flow rate, cell heatertemperature, suppressor current, etc.) depend on theapplication to be run. Refer to the column manual forthe required parameters for your application.Set or verify system operating parameters from the Chromeleon Control Panel.Operating parameters can also be set automatically by loading a Chromeleonsequence.NOTEClicking the <strong>System</strong> Startup button on the ChromeleonControl Panel, starts the pump and suppressor. The flowrate and suppressor current settings that were in effectwhen the system was shut down are restored.1. Verify that the pump is on and set to the correct flow rate.2. Verify that the suppressor current is on and that the setting is correct.3. Verify that the cell heater is set to the correct value.4. Verify that the column heater is set to the correct value.3-10 Doc. 031879-01 3/03