<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>4. Unscrew the sixthumbscrews on thetop metal plate (seeFigure B-18) andremove the plate.5. Go on toSection B.9.2.Thumbscrew (6)Figure B-18. <strong>ICS</strong>-<strong>1000</strong> Column Heater Cover RemovedB.9.2Installing the Columns1. Remove the union connecting GUARD IN and GUARD OUT.2. Connect GUARD IN to the guard column inlet. Check the arrow on thecolumn label; it should point down (flow away from the injectionvalve).3. Connect GUARD OUT to the guard column outlet.4. Push the guard column onto the back column clip.5. Remove the union connecting COL IN and COL OUT and6. Connect COL IN to the separator column inlet. Connect COL OUT tothe separator column outlet. Check the arrow on the column label. Itshould point up (flow toward the cell).7. Push the separator column onto the front column clips.8. Replace the top metal plate on the column heater (if installed).B-20 Doc. 031879-01 3/03