<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>device icon under the timebase, and select Properties from the Editmenu. Select the Options tab (see Figure 2-10).Figure 2-10. Server Configuration Properties: Degas Mode OptionsDegas Mode Options• Always Off: The degas pump is always off.• Always On: The degas pump is always on. This setting is for testpurposes by a Dionex Service Representative. Do not use this settingfor routine operation.• Cycle: The degas pump cycles on and off according to the timesspecified in the Cycle On and Off fields. Cycle On specifies for howlong the degas pump runs during a cycle. Cycle Off specifies the timebetween cycles.• Monitor: (default mode) The <strong>ICS</strong>-<strong>1000</strong> monitors the degas pressurereading and turns the degas pump on and off as required.2-20 Doc. 031879-01 3/03