<strong>ICS</strong>-<strong>1000</strong> <strong>Ion</strong> <strong>Chromatography</strong> <strong>System</strong>SuppressorThe suppressor reduces the eluent conductivity and enhances theconductivity of the sample ions, thereby increasing detection sensitivity.Either an AES® Atlas Electrolytic Suppressor, SRS® Self-RegeneratingSuppressor, or MMS MicroMembrane Suppressor can be used withthe <strong>ICS</strong>-<strong>1000</strong>. See Section 2.4.6 for details.DS6 Heated Conductivity CellThe flow-through conductivity cell measures the electrical conductanceof analyte ions as they pass through the cell. A heat exchanger inside thecell regulates the temperature, which can be set to between 30 °C and55 °C. For optimum performance, set the temperature to at least 7 °Cabove the ambient temperature. See Section 2.4.7 for details about thecell.Tubing ChasesThe upper tubing chase routes tubing from the top cover to the componentpanel. The lower tubing chase routes tubing from the component panelthrough the <strong>ICS</strong>-<strong>1000</strong> interior to the rear panel.2-6 Doc. 031879-01 3/03