Installing and Plumbing the Columns and SuppressorB.9.3Installing the Suppressor1. Locate the suppressormounting tabs on thecomponent panel (seeFigure B-19). Reachinto the openingbetween the tabs andpull the yellowconnector and attachedcable slightly out of theopening.2. Connect this cable to thecable from thesuppressor (align thepins on the twoconnectors and pushthem together).4311 211 SuppressorMounting Tabs2 Suppressor CableConnector3 Guideline for AES4 Guideline for SRSFigure B-19. Suppressor Mounting Tabsand Guidelines3. Orient the suppressor with the REGEN OUT fitting on top and thecable to the right.4. Feed the cable into the opening in the component panel.5. Align the top of the suppressor with the guideline printed on thecomponent panel (see Figure B-19) and align the slots on the rear ofthe suppressor with the tabs on the panel.Doc. 031879-01 3/03 B-21