3 • Operation and Maintenancestand-alone operation. This requires a TTL cable connection between the AS50and the <strong>ICS</strong>-<strong>1000</strong>. See Section C.3 for stand-alone setup instructions.3.10.1 Loading Samples with a Syringe1. Make sure the injection port on the <strong>ICS</strong>-<strong>1000</strong> front door (seeFigure 2-1) is connected to the sample port S (5) on the injection valve(see Figure 3-9).Figure 3-9. Injection Valve Connections2. Fill the 1 cc syringe (P/N 016388) provided in the <strong>ICS</strong>-<strong>1000</strong> Ship Kit(P/N 057905) with a calibration standard or sample.3. Insert the syringe into the injection port on the <strong>ICS</strong>-<strong>1000</strong> front door.4. Verify that the injection valve is in the Load position.5. Overfill the sample loop with several sample loop volumes. Excesssample will exit through the injection valve waste line.6. Leave the syringe in the port.7. Switch the injection valve to the Inject position (see Section 3.10.4).Doc. 031879-01 3/03 3-15