Respiration Monitoring Respiration and CO 2 Troubleshooting<br />

5.4 Respiration and CO 2 Troubleshooting<br />

MESSAGE/PROBLEM REASON SOLUTION<br />

Resp. Waveform Too Large Scales set inappropriately. Change lead selection.<br />

Change Respiration scale.<br />

Resp. Waveform Too Small Patient breathing is shallow<br />

or patient is turned on side.<br />

Change lead selection.<br />

Scale set inappropriately. Change respiration scale.<br />

False Apnea Alarm Apnea delay may be Choose another apnea<br />

improperly set.<br />

delay<br />

Patient may be having Reposition electrodes to<br />

frequent episodes of CVA. better detect respirations.<br />

Scale size may be too low. Change Respiration scale<br />

No Resp. Waveform or Respiration turned Off. Turn respiration On<br />

Rate Displayed<br />

(Off will be displayed in<br />

Resp. tile).<br />

Patient connected using<br />

ESIS choke cable.<br />

Cable not connected.<br />

CO2 not selected in<br />

Display Setup Menu.<br />

“CHK Lead” Message Increased impedance<br />

caused by one of the<br />

following:<br />

Chest hair under<br />

electrodes.<br />

Check that proper patient<br />

cable is used. Use approved<br />

non-ESIS patient cable.<br />

Check cable.<br />

Set CO2 in Display Setup<br />

Menu.<br />

Prep chest.<br />

Dried electrode gel. Change electrodes.<br />

Electrode off. Replace electrode.<br />

Lead off. Replace lead.<br />

Cracked lead wires. Replace lead wires.<br />

Poor skin prep. Clean and abrade skin<br />

before applying electrodes.<br />

“CVA” Message Can be caused by shallow<br />

breathing or an apnea<br />

event.<br />

Check the patient.<br />

Patient HR and respiratory<br />

rate identical.<br />

Adjust scales or leads if<br />

necessary.<br />

Passport V Operating Instructions 0070-10-0704-02 5 - 5