Appendix Safety Designations<br />

19.14 Safety Designations<br />

19.14.1 Safety Classification<br />

Transportable equipment intended for mobility<br />

Type of Protection Against Electric Shock<br />

Class I equipment<br />

Where the integrity of the external protective earth conductor arrangement is in doubt,<br />

equipment shall be operated from its internal electrical electric power source (battery).<br />

Degree of Protection Against Electric Shock<br />

ECG/RESP/IBP/TEMP: CF, defibrillation-proof<br />

SpO2/NIBP/CO2: BF, defibrillation-proof<br />

Mode of Operation<br />

Continuous<br />

Protection Against Hazards of Explosion<br />

Not protected (ordinary equipment)<br />

Not for use in the presence of flammable anesthetic mixture, O2 or N2O. Protection Against Ingress of Liquids<br />

Complies with the requirements of IEC60529: 2001, IPX1.<br />

Complies with the requirements of IEC60601-1:1988+A1:1991+A2:1995.<br />

Electrical Connection Between Equipment and Patient<br />

Direct electrical or non-electrical connection with the patient<br />

19.14.2 Safety Performance Index<br />

Shock<br />

Complies with the requirements specified in ISO 9919: 2005, clause 21.101.<br />

Vibration<br />

Complies with the requirements of ISO9919: 2005, clause 21.101 b).<br />

Complies with the requirements of IEC60068-2-6:1995, sinusoidal vibration.<br />

Drop<br />

Main Unit Drop<br />

Complies with the requirements of ECRI PB-296 892:1979, section AIII 3.3.<br />

Packaging Drop<br />

Complies with the requirements specified in ISTA 2A:2006.<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 31