CO 2 Monitoring (optional) CO 2 Menu<br />

CO 2 Menu<br />

MENU ITEM SELECTIONS COMMENTS<br />

Selection Gas Module 3<br />

Status<br />

10.3.1 CO 2 Setup Menu (DPM Sidestream Only)<br />

To display the CO 2 Setup Menu:<br />

Internal CO 2<br />

Module Status<br />

1. On the front panel:<br />

Press the Normal Screen key to return to the normal screen.<br />

On the normal screen:<br />

Select Parameters 〉 CO2 〉 CO2 Setup<br />

or<br />

Select the CO2 tile 〉 CO2 Setup.<br />

NOTE: To extend the lifetime of the watertrap and module,<br />

disconnect the watertrap and set the operating mode to<br />

Standby mode when CO 2 monitoring is not required.<br />

NOTE: The watertrap collects water drops condensed in the<br />

sampling line and therefore prevents them from entering<br />

the module. It is important to periodically drain the<br />

watertrap to avoid water blocking the airway.<br />

Active CO 2<br />

Source<br />

Gas Module OFF OFF None<br />

Gas Module OFF ON Internal CO2 Module<br />

Gas Module ON OFF Gas Module<br />

Gas Module ON ON Gas Module<br />

Internal CO2 Module<br />

OFF OFF None<br />

Internal CO 2<br />

Module<br />

Internal CO 2<br />

Module<br />

Internal CO 2<br />

Module<br />

OFF ON Internal CO 2 Module<br />

ON OFF Gas Module<br />

ON ON Internal CO 2 Module<br />

Note: ON/OFF Status of the Module:<br />

If the Gas Module is connected to a serial port configured to ‘Gas Module’, and<br />

the power is switched on, the module is considered to be ON. Otherwise it is<br />

considered to be OFF.<br />

If the Oridion CO 2 Module is installed and a Filterline is connected, the module<br />

is considered to be ON. Otherwise it is considered to be OFF.<br />

If the DPM CO 2 Module and water trap are installed, the module is considered<br />

to be ON, Otherwise it is considered to be OFF.<br />

Flow Rate 70 ml/min<br />

100 ml/min (default)<br />

120 ml/min<br />

150 ml/min<br />

Applicable to DPM Sidestream CO 2<br />

module only.<br />

Note: For Oridion Microstream CO 2, the<br />

flow rate is 50ml/min. It is not user<br />

adjustable.<br />

Passport V Operating Instructions 0070-10-0704-02 10 - 5