NIBP Monitoring NIBP Menu<br />

7.3 NIBP Menu<br />

To display the NIBP Menu:<br />

1. On the front panel:<br />

Press the Normal Screen key to return to the normal screen.<br />

2. On the normal screen:<br />

Select Parameters 〉 NIBP<br />

or<br />

Select the NIBP tile<br />

FIGURE 7-3 NIBP Menu<br />

NIBP Menu<br />

MENU ITEM SELECTIONS COMMENTS<br />

Normal Screen — Select to return to normal screen (or press the<br />

Normal Screen key on the front panel).<br />

Set Start Pressure Adult: 100 to 280<br />

(default = 180)<br />

Pediatric: 60 to 180<br />

(default = 140)<br />

Neonate: 40 to 120<br />

(default = 100)<br />

Select to adjust initial NIBP pump up pressure in<br />

increments of 5. The selection range varies<br />

according to the Patient Size.<br />

The Set Start Pressure cannot be changed during<br />

an NIBP measurement or the five minutes STAT<br />

measurement.<br />

Passport V Operating Instructions 0070-10-0704-02 7 - 5