Performance Characteristics Appendix<br />

miniMediCO 2 Microstream ®<br />

Range: 0 - 99 mmHg<br />

Accuracy specification of the measured CO2 partial pressure is according to the<br />

following table (This testing is done according to ISO 21647 clauses 51.101.1 and<br />

51.101.2.).<br />

CO2 PARTIAL PRESSURE * ACCURACY**<br />

0 - 38 mmHg ± 2 mmHg<br />

39 - 99 mmHg ± (5% of reading + 0.08% for every 1 mmHg above 38 mmHg)<br />

* At sea level.<br />

** Accuracy applies for respiration rates of up to 80 RPM. For respiration rates above 80 RPM, accuracy is<br />

4mmHg or ± 12% of the reading, whichever is greater, for EtCO2 values exceeding 18 mmHg. This is<br />

tested according to and is compliant with ISO 21647. To achieve the specified accuracies for respiration<br />

rates above 60 respirations/minute, the Microstream FilterLine H Set for Infant/Neonatal<br />

(p/n 006324) must be used. Above 55 degrees C module temperature, ± 1 mmHg or ± 2.5% (whichever is<br />

greater) has to be added to the tolerance of the accuracy specs.<br />

Accuracy in the presence of interfering gases:<br />

The accuracy specification is maintained to within 4% of the values indicated in the table<br />

above in the presence of interfering gases according to ISO 21647 clauses 51.101.3 and<br />

101.1. The accuracy specification is maintained to within 4% of the values indicated in the<br />

table above in the presence of up to 80% Helium with up to 15% oxygen when tested<br />

according to the procedures defined by ISO 21647.<br />

Flow Rate: 50 ml/min -7.5 + 15 ml/min, flow measured by volume.<br />

Respiration Rate Range: 0 - 150 RPM<br />

Respiration Rate Accuracy:<br />

Respiration Rate 0 to 70 RPM 71 to 120 RPM 121 to 150 RPM<br />

Accuracy ± 1 RPM ± 2 RPM ± 3 RPM<br />

Self Maintenance (SFM) Interval:<br />

Self-Maintenance (SFM) is performed only during measurement mode. The module performs<br />

one or more of the following:<br />

Ambient pressure measurement<br />

Auto zero (AZ)<br />

Flow test<br />

SFM is triggered under the following conditions:<br />

During the first hour after entering measurement mode, periodically for durations of<br />

typically 10 seconds at a rate which limits the total time consumed by SFMs to less<br />

than 2% of the time in which active measurements are taken. Following the first hour<br />

19 - 22 0070-10-0704-02 Passport V Operating Instructions