SpO2 Monitoring SpO 2 Pulse Oximetry<br />
6.4.2 Nellcor ® SpO 2<br />
CAUTION: When equipped with Nellcor® SpO2, use only Nellcor<br />
oxygen transducers including Nellcor Oxisensor® and<br />
OxiMax® patient dedicated adhesive sensors. Use of other<br />
oxygen transducers may cause improper oximeter<br />
performance.<br />
Nellcor Sensors<br />
Nellcor provides a family of sensors suitable for a wide variety of clinical settings and<br />
patients. Specific sensors have been developed for neonates, infants, children, and adults.<br />
Oxisensor ® and OxiMax ® oxygen transducers are sterile adhesive sensors with optical<br />
components mounted on adhesive tape. Oxiband ® oxygen transducers and the Duraform ®<br />
oxygen transducer system are reusable sensors that are applied with disposable adhesive.<br />
The Durasensor DS-100A Adult Digit Oxygen Transducer is a reusable sensor with its optical<br />
components mounted in a plastic casing. The Nellcor RS-10 and Max-Fast ® oxygen<br />
transducers are adhesive sensors for application on the forehead or temples.<br />
Packaging of sterile sensors shall ensure sterile conditions until opened or damaged or until<br />
expiration date is reached.<br />
NOTE: Consideration should be given to the disposal of packaging<br />
waste.<br />
To order all Nellcor accessories and sensors call 1-888-744-1414.<br />
Selecting a Nellcor Sensor<br />
Sensors are designed for specific sites on patients with designated weight ranges. To select<br />
the appropriate sensor, consider the patient’s weight, level of activity, adequacy of perfusion,<br />
which sensor sites are available, whether sterility is required, and the anticipated duration of<br />
monitoring.<br />
NOTE: Only Nellcor oxygen transducers should be used with the<br />
Passport V monitors with Nellcor OxiMax ® pulse oximetry.<br />
Sat-Seconds Alarm Management<br />
With traditional alarm management, high and low alarm limits are set for monitoring oxygen<br />
saturation. During monitoring, as soon as an alarm limit is violated, an audible alarm<br />
immediately sounds. When the patient % SpO2 fluctuates near an alarm limit, the alarm<br />
sounds each time the limit is violated. Such frequent alarm can be distracting. Nellcor’s Sat-<br />
Seconds alarm management technique is used to reduce these nuisance alarms.<br />
The Sat-Seconds feature is available with the Nellcor SpO2 module to decrease the<br />
likelihood of false alarms caused by motion artifacts. To set the Sat-Seconds limit, select Sat-<br />
Seconds in the SpO2 Setup menu and then select the appropriate setting.<br />
Passport V Operating Instructions 0070-10-0704-02 6 - 7