SpO2 Monitoring SpO 2 Pulse Oximetry<br />

6.4.2 Nellcor ® SpO 2<br />

CAUTION: When equipped with Nellcor® SpO2, use only Nellcor<br />

oxygen transducers including Nellcor Oxisensor® and<br />

OxiMax® patient dedicated adhesive sensors. Use of other<br />

oxygen transducers may cause improper oximeter<br />

performance.<br />

Nellcor Sensors<br />

Nellcor provides a family of sensors suitable for a wide variety of clinical settings and<br />

patients. Specific sensors have been developed for neonates, infants, children, and adults.<br />

Oxisensor ® and OxiMax ® oxygen transducers are sterile adhesive sensors with optical<br />

components mounted on adhesive tape. Oxiband ® oxygen transducers and the Duraform ®<br />

oxygen transducer system are reusable sensors that are applied with disposable adhesive.<br />

The Durasensor DS-100A Adult Digit Oxygen Transducer is a reusable sensor with its optical<br />

components mounted in a plastic casing. The Nellcor RS-10 and Max-Fast ® oxygen<br />

transducers are adhesive sensors for application on the forehead or temples.<br />

Packaging of sterile sensors shall ensure sterile conditions until opened or damaged or until<br />

expiration date is reached.<br />

NOTE: Consideration should be given to the disposal of packaging<br />

waste.<br />

To order all Nellcor accessories and sensors call 1-888-744-1414.<br />

Selecting a Nellcor Sensor<br />

Sensors are designed for specific sites on patients with designated weight ranges. To select<br />

the appropriate sensor, consider the patient’s weight, level of activity, adequacy of perfusion,<br />

which sensor sites are available, whether sterility is required, and the anticipated duration of<br />

monitoring.<br />

NOTE: Only Nellcor oxygen transducers should be used with the<br />

Passport V monitors with Nellcor OxiMax ® pulse oximetry.<br />

Sat-Seconds Alarm Management<br />

With traditional alarm management, high and low alarm limits are set for monitoring oxygen<br />

saturation. During monitoring, as soon as an alarm limit is violated, an audible alarm<br />

immediately sounds. When the patient % SpO2 fluctuates near an alarm limit, the alarm<br />

sounds each time the limit is violated. Such frequent alarm can be distracting. Nellcor’s Sat-<br />

Seconds alarm management technique is used to reduce these nuisance alarms.<br />

The Sat-Seconds feature is available with the Nellcor SpO2 module to decrease the<br />

likelihood of false alarms caused by motion artifacts. To set the Sat-Seconds limit, select Sat-<br />

Seconds in the SpO2 Setup menu and then select the appropriate setting.<br />

Passport V Operating Instructions 0070-10-0704-02 6 - 7