Menus: ECG Main and Submenus ECG Monitoring<br />

Arrhythmia Menu<br />

4.4.3 ST Menu<br />

MENU ITEM SELECTIONS COMMENTS<br />

Asystole Delay 3 to 10 seconds<br />

default = 5<br />

V-Tach Rate 100 to 180 BPM<br />

default = 130<br />

V-Tach PVC 3 to 15<br />

default = 3<br />

FIGURE 4-8 ST Menu<br />

Select to change the Asystole Delay Time.<br />

Select to change the V-Tach alarm rate limit.<br />

Select to change the required number of PVCs detected<br />

in a continuous sequence before V-Tach is detected and<br />

labeled.<br />

NOTE: The ST Menu is disabled if Patient Size = Neonate.<br />

4 - 10 0070-10-0704-02 Passport V Operating Instructions