Messages Prompt Messages<br />

14.3.10 Power Supply<br />

Keyboard Initialization Error Contact Technical Support.<br />

Keyboard SelfTest Error Contact Technical Support.<br />

Defib Sync signal error Contact Technical Support.<br />

CF card format complete CF card format complete<br />

successfully.<br />

CF card format error CF card occur error. Contact Technical Support.<br />

ALARM MESSAGE REASON SOLUTION<br />

12V Too High The monitoring of the internal<br />

voltage finds that 12V voltage is<br />

too high.<br />

Contact Technical Support.<br />

12V Too Low The monitoring of the internal<br />

voltage finds that 12V voltage is<br />

too low.<br />

5V Too High The monitoring of the internal<br />

voltage finds that 5V voltage is<br />

too high.<br />

5V Too Low The monitoring of the internal<br />

voltage finds that 5V voltage is<br />

too low.<br />

3.3V Too High The monitoring of the internal<br />

voltage finds that 3.3V voltage is<br />

too high.<br />

3.3V Too Low The monitoring of the internal<br />

voltage finds that 3.3V voltage is<br />

too low.<br />

14.3.11 Data Management<br />

Prompt messages Concerning Patient Data Transfer<br />

Exported Logs<br />

Contact Technical Support.<br />

Contact Technical Support.<br />

Contact Technical Support.<br />

Contact Technical Support.<br />

Contact Technical Support.<br />

MESSAGE REASON SOLUTION<br />

Patient data transfer error. Patient data transfer error. Check the storage device and<br />

retransfer patient data.<br />

Downloading Patient Data Remain in the state of patient<br />

data downloading.<br />

No handling needed.<br />

Uploading Patient Data Remain in the state of patient<br />

data uploading.<br />

Patient data download<br />

complete.<br />

Patient data upload<br />

complete.<br />

Patient data downloading<br />

finishes.<br />

No handling needed.<br />

No handling needed.<br />

Patient data uploading finishes. No handling needed.<br />

Passport V Operating Instructions 0070-10-0704-02 14 - 19