ECG Monitoring Preparation and Lead Placement<br />

4.5.3.5 Lead Placement: Lead II Monitoring<br />

The recommended lead placement for Lead II monitoring is as follows.<br />

White<br />

RA LA<br />

FIGURE 4-15 Lead II Monitoring (AHA) FIGURE 4-16 Lead II Monitoring (IEC)<br />

Place the RA (white) electrode under the<br />

patient’s right clavicle, at the midclavicular<br />

line within the rib cage frame.<br />

Place the LA (black) electrode under the<br />

patient’s left clavicle, at the midclavicular<br />

line within the rib cage frame.<br />

Place the LL (red) electrode on the<br />

patient’s lower left abdomen within the<br />

rib cage frame.<br />

Select ECG Lead II on the monitor. Lead II is<br />

the direct electrical line between the RA<br />

(white) electrode and the LL (red) electrode.<br />

LL<br />

Black<br />

Red<br />

R L<br />

Place the R (red) electrode under the<br />

patient’s right clavicle, at the midclavicular<br />

line within the rib cage frame.<br />

Place the L (yellow) electrode under the<br />

patient’s left clavicle, at the midclavicular<br />

line within the rib cage frame.<br />

Place the F (green) electrode on the<br />

patient’s lower left abdomen within the<br />

rib cage frame.<br />

Select ECG Lead II on the monitor. Lead II is<br />

the direct electrical line between the R (red)<br />

electrode and the F (green) electrode.<br />

Passport V Operating Instructions 0070-10-0704-02 4 - 19<br />

Red<br />

F<br />

Yellow<br />

Green