Prompt Messages Messages<br />

NIBP: STAT Displayed while the Interval is<br />

set to STAT.<br />

Press START to initiate a STAT<br />

measurement sequence or use<br />

the ‘Interval’ key to change the<br />

setting to Off or a timed<br />

interval value.<br />

NIBP: Warming Up Warming Up This is normal operation. Wait<br />

for message to clear.<br />

NIBP: Retry Pump Higher A measurement has been<br />

attempted but no reading was<br />

possible. This results from<br />

inadequate cuff inflation<br />

pressure.<br />

NIBP: Retry Cuff is loosely wrapped. The cuff<br />

may be too loosely wrapped or<br />

not attached at all.<br />

Weak signal. The pulse of the<br />

patient may be too weak or the<br />

cuff is loosely wrapped.<br />

Excessive motion. In<br />

measurement, signals contain<br />

motion artifact or too much<br />

interference.<br />

Timeout. Measurement takes<br />

more than 120 seconds in ADU/<br />

PED mode and 90 seconds in<br />

NEO mode.<br />

Retry Overpressure or bad<br />

measurement.<br />

Accuracy Testing... Appears when press Accuracy<br />

Test button in Calibrate NIBP<br />

Menu.<br />

Accuracy Testing Complete. During the accuracy testing,<br />

press Stop Accuracy Test button<br />

in Calibrate NIBP Menu, this<br />

message will display.<br />

Leakage Testing... Appears when press Leakage<br />

Test button in Calibrate NIBP<br />

Menu.<br />

Leakage Testing Complete During the leakage testing, press<br />

Stop Leakage Test button in<br />

Calibrate NIBP Menu, this<br />

message will display.<br />

Calibrating... Occurs when NIBP module is<br />

calibrating.<br />

Calibration Failed During calibration, departure<br />

occurs between the<br />

measurement pressure and<br />

reference pressure due to<br />

improper operation or<br />

measurement error.<br />

Retry will be attempted.<br />

Check that appropriate patient<br />

size is set.<br />

Preset initial inflation<br />

pressure.<br />

Retry will be attempted. Check<br />

for leaks and quality of<br />

peripheral pulses. Decrease<br />

patient movement. Switch cuff<br />

to another limb.<br />

14 - 12 0070-10-0704-02 Passport V Operating Instructions<br />

/<br />

/<br />

/<br />

/<br />

Wait until calibration is<br />

completed<br />

Repeat calibration<br />

procedure.