Prompt Messages Messages<br />
NIBP: STAT Displayed while the Interval is<br />
set to STAT.<br />
Press START to initiate a STAT<br />
measurement sequence or use<br />
the ‘Interval’ key to change the<br />
setting to Off or a timed<br />
interval value.<br />
NIBP: Warming Up Warming Up This is normal operation. Wait<br />
for message to clear.<br />
NIBP: Retry Pump Higher A measurement has been<br />
attempted but no reading was<br />
possible. This results from<br />
inadequate cuff inflation<br />
pressure.<br />
NIBP: Retry Cuff is loosely wrapped. The cuff<br />
may be too loosely wrapped or<br />
not attached at all.<br />
Weak signal. The pulse of the<br />
patient may be too weak or the<br />
cuff is loosely wrapped.<br />
Excessive motion. In<br />
measurement, signals contain<br />
motion artifact or too much<br />
interference.<br />
Timeout. Measurement takes<br />
more than 120 seconds in ADU/<br />
PED mode and 90 seconds in<br />
NEO mode.<br />
Retry Overpressure or bad<br />
measurement.<br />
Accuracy Testing... Appears when press Accuracy<br />
Test button in Calibrate NIBP<br />
Menu.<br />
Accuracy Testing Complete. During the accuracy testing,<br />
press Stop Accuracy Test button<br />
in Calibrate NIBP Menu, this<br />
message will display.<br />
Leakage Testing... Appears when press Leakage<br />
Test button in Calibrate NIBP<br />
Menu.<br />
Leakage Testing Complete During the leakage testing, press<br />
Stop Leakage Test button in<br />
Calibrate NIBP Menu, this<br />
message will display.<br />
Calibrating... Occurs when NIBP module is<br />
calibrating.<br />
Calibration Failed During calibration, departure<br />
occurs between the<br />
measurement pressure and<br />
reference pressure due to<br />
improper operation or<br />
measurement error.<br />
Retry will be attempted.<br />
Check that appropriate patient<br />
size is set.<br />
Preset initial inflation<br />
pressure.<br />
Retry will be attempted. Check<br />
for leaks and quality of<br />
peripheral pulses. Decrease<br />
patient movement. Switch cuff<br />
to another limb.<br />
14 - 12 0070-10-0704-02 Passport V Operating Instructions<br />
/<br />
/<br />
/<br />
/<br />
Wait until calibration is<br />
completed<br />
Repeat calibration<br />
procedure.