Appendix Performance Characteristics<br />

Rise times (t10–90%) 3<br />

@120 ml/min<br />

Delay time 4 (t0—10%) < 4 s<br />

Identification Dual agent<br />

CO2 250 ms (fall time 200 ms)<br />

N2O 250 ms<br />

O2 600 ms<br />

HAL, ISO, SEV, DES 300 ms<br />

ENF 350 ms<br />

Primary agent ID threshold 5 0.15% (0.4% during ISO accuracy mode)<br />

Secondary agent ID threshold 5 0.3% (0.5% during ISO accuracy mode) or 5% REL<br />

(10% REL for Isoflurane) of primary agent if primary<br />

agent >10%<br />

Agent ID time Three breaths - Typically less than 10 s<br />

Display Update Rate Breath-by-Breath<br />

Compensation The Gas Module 3 is equipped with automatic barometric<br />

pressure compensation.<br />

Main Fuse 2x T0.8A 250V<br />

1 Includes interference from other gases.<br />

2 The step rise time specification at 200 ml/min sample flow includes DRYLINE• Water Trap, Adult/Pediatric<br />

and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m.<br />

3 The step rise time specification at 120 ml/min sample flow includes DRYLINE• Water Trap, Neonate and<br />

DRYLINE• Sampling Line, Neonate 2.5 m.<br />

4 The delay time specification is valid both for 120 ml/min sample flow (using DRYLINE• Water Trap, Neonate<br />

and DRYLINE• Sampling Line, Neonate 2.5 m) and for 200 ml/min sample flow (using DRYLINE•<br />

Water Trap, Adult/Pediatric and DRYLINE• Sampling Line, Adult/Pediatric 2.5 m).<br />

5 For HAL, add 0.1% ABS to threshold values.<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 27