Appendix Safety Designations<br />

Dielectric Strength<br />

Complies with the requirements of IEC 60601-1: 1997, clause 20.<br />

AC power to earth(A-a1): 1500VRMS, 1min/5mA<br />

AC power to applied part (B-a): 4000VRMS, 1min/5mA<br />

Applied part to earth(B-d): 1500VRMS, 1min/5mA<br />

Between applied parts(B-b): 1500VRMS, 1min/5mA<br />

Earthing Impedance<br />

Complies with the requirements of IEC 60601-1: 1997, clause 18.<br />

The impedance between the protective earth terminal of the power input connector and<br />

any tangible metal part (such as screw and equipotential stud) that is protectively earthed<br />

shall not exceed 0.1 ohms.<br />

Protective Earthing and Potential Equalization<br />

Complies with the requirements of IEC 60601-1: 1997, clause 58.<br />

The protective earth terminal cannot be used for mechanical connections between different<br />

parts of the equipment or for fixing any component irrelevant to protective or functional<br />

earthing.<br />

Anti-power Interference<br />

After interference from ESU/defibrillation/electrostatic discharge<br />

The equipment returns to normal operation mode within 10 seconds. The display is able to<br />

show physiological parameters and waveforms normally. For example, the ECG waveform<br />

data stored are not lost.<br />

Protection Against Power Failure<br />

Requirements of ISO21647: 2005: provides the device to prevent AC power from<br />

disconnection and tension of 10 Kg.<br />

If the power supply is restored after interrupted, saved presets, user settings, and patient data<br />

previous to power failure are retained or can be restored with a DPM storage device.<br />

The monitor stores up to 96 hours of trend information (approximately 6000 trend items)<br />

when powered off. The interval for trend data is 1 minute.<br />

Power Input<br />

Complies with the requirements of IEC60601-1: 1997, clause 7.1.<br />

Limitation of Voltage and (or) Energy<br />

Complies with the requirements of IEC60601-1: 1997, clause 1.5.<br />

Defibrillation-proof Applied Part<br />

Complies with the requirements of IEC60601-1: 1997, clause 17.h.<br />

Passport V Operating Instructions 0070-10-0704-02 19 - 33