Prompt Messages Messages<br />

SpO 2 : Interference Patient Movement.<br />

External interference.<br />

Noise detected on<br />

the pulse signal<br />

prevents pulse<br />

discrimination.<br />

14.3.4 Temp Prompt Message<br />

Decrease patient<br />

motion,check sensor.<br />

Contact Technical<br />

Support.<br />

Masimo/Nellcor<br />

SpO2: No Sensor No sensor Apply sensor Masimo/DPM/<br />

Nellcor<br />

SpO2: Weak Pulse Low pulse amplitude Check the condition of<br />

poor perfusion for<br />

patient.<br />

Contact Technical<br />

Support.<br />

Nellcor<br />

SpO 2: Low Signal The SpO 2 signal is<br />

too low or too weak.<br />

SpO 2: Weak Signal The SpO 2 signal is<br />

too low or too weak.<br />

SpO 2: Check Sensor Possible Faulty<br />

sensor<br />

Check sensor<br />

placement, move as<br />

necessary.<br />

Switch limb / Notify<br />

physician.<br />

Check patient<br />

Reposition sensor<br />

Change sensor<br />

Nellcor<br />

Masimo<br />

Replace Sensor Nellcor<br />

SpO 2: Motion Patient movement Check patient Nellcor<br />

SpO 2 : Pulse Search Monitor searching<br />

for pulse.<br />

Sensor just applied<br />

Poor perfusion at<br />

sensor site.<br />

Check patient.<br />

Reposition sensor.<br />

Masimo/DPM/<br />

Nellcor<br />

ALARM MESSAGE REASON SOLUTION<br />

Temp: Communication Stop As the Temp module<br />

communication stops, the data<br />

packets sent by the Temp module<br />

can not be received.<br />

Contact Technical Support.<br />

Temp: SelfTest Error Module send out error during<br />

selftest because of some<br />

unspecified reasons.<br />

Contact Technical Support.<br />

Temp: Calibration Error A calibration failed. Restart the monitor.<br />

Contact Technical Support.<br />

Temp: Initialization Error During the Temperature module<br />

power-on,as the Temperature<br />

module communication<br />

stops,system fails to<br />

communicate with module.<br />

Contact Technical Support.<br />

14 - 10 0070-10-0704-02 Passport V Operating Instructions