ECG Monitoring Arrhythmia Alarms (optional)<br />

4.7.2.4 Irregular Heart Rate Alarm<br />

The Irregular Heart Rate alarm is activated when the measured variations in the R-R<br />

interval over a period of time exceeds a preset limit established by the arrhythmia algorithm.<br />

The Irregular Heart Rate alarm is an alarm event that produces:<br />

4.7.2.5 PVC/minute Alarm<br />

4.7.2.6 Run Alarm<br />

Alarm visual and audio alarm indicators.<br />

An IRREGULAR HR text message above the ECG1 waveform area.<br />

The High PVC alarm is activated when the number of PVCs detected per minute exceeds the<br />

configured threshold. The PVC limit can be set to Off, or 1 to 30 PVCs per minute.<br />

The High PVC alarm is an alarm event that produces :<br />

Alarm visual and audio alarm indicators.<br />

A High PVC text message above the ECG1 waveform area.<br />

NOTE: PVC/min will not be displayed during periods of Ventricular<br />

Rhythms, V-TACH, V-FIB and Asystole.<br />

The Run alarm is activated when the number of consecutive PVCs occur at a rate that equals<br />

or exceeds the user defined V-Tach Rate. The number of consecutive PVCs that constitute a<br />

Run is one beat less than the minimum used to identify V-Tach.<br />

The Run alarm is an alarm event that produces:<br />

4.7.2.7 Trigeminy Alarm<br />

Alarm visual and audio alarm indicators.<br />

A RUN text message above the ECG1 waveform area.<br />

The Trigeminy alarm is activated when three or more cycles of one PVC coupled to two<br />

normal beats are detected. This rhythm could also cause an Irregular HR alarm.<br />

The Trigeminy alarm is an alarm event that produces:<br />

Alarm visual and audio alarm indicators.<br />

A TRIGEMINY text message above the ECG1 waveform area.<br />

4.7.2.8 Ventricular Rhythm (V-Rhythm) Alarm<br />

The V-Rhythm alarm is activated when more than 2 consecutive PVCs occur at a rate that is<br />

less than the user defined V-Tach Rate.<br />

The V-Rhythm alarm is an alarm event that produces:<br />

Alarm visual and audio alarm indicators.<br />

A VENTRICULAR RHYTHM text message above the ECG1 waveform area.<br />

Passport V Operating Instructions 0070-10-0704-02 4 - 27